Functional description
Display backlight
Press the [!] key to turn the display backlight on or off. The backlight
is switched off automatically during the measurement.
Recall of stored data
If the instrument is switched on, press the [!] key for more than
4 seconds to access the recall menu.
!
!
!
AbS
Measuring the absorbance
This Mode function allows measurement of the absorbance at the
specified wavelength. The result appears in the display in mAbs.
1000 mAbs = 1 A (Absorbance unit)
Abs1:
= 528 nm
Abs2:
= 605 nm
Countdown / reaction period
If a reaction period is included in a method a countdown function
can be used:
Press the [!] key and hold.
Press the [ZERO/TEST] key.
Release the [!] key; the countdown starts.
After the countdown is finished the measurement starts automati-
cally.
It is possible to interrupt the countdown by pressing the [ZERO/TEST]
key. Measurement starts immediately.
Caution:
An incomplete reaction period can lead to incorrect test
results.
Zero
Test

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lovibond CheckitDirect and is the answer not in the manual?

Lovibond CheckitDirect Specifications

General IconGeneral
BrandLovibond
ModelCheckitDirect
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals