Technical data – Operating messages
Operating messages
Measuring range exceeded or excessive turbidity.
Result below the lowest limit of the measuring range.
Replace 9 V battery, no further tests possible.
Battery capacity is too low for the display backlight;
measurement is still possible.
Technical Data
Light source: LED, Filter (
1
= 528 nm)
LED (
2
= 605 nm)
Battery: 9 V-block battery (Life approx. 600 tests, without display light)
Auto-OFF: Automatic switch off 7 minutes after last keypress
Ambient conditions: 5–40°C
30–90% rel. humidity (non-condensing)
CE: Certificate for Declaration of CE-Conformity
at www.tintometer.com
To ensure maximum accuracy of test results, always use the reagent systems
supplied by the instrument manufacturer.