GB
Technical Data
Technical Data
Instrument triple wavelength, automatic wavelength selection,
direct reading colorimeter
Light source: LEDs, interference filters (IF) and photosensor in transparent
cell chamber. Wavelength specifications of the IF:
530 nm
= 5 nm
560 nm
= 5 nm
610 nm = 6 nm
Wavelength accuracy ± 1 nm
Photometric accuracy* 3% FS (T = 20° C – 25° C)
Photometric resolution 0.01 A
Power supply 4 batteries (AA/LR6)
Operating time
53hr operating time or 15000 test measurements in continuous
mode when display backlight is off
Auto-OFF automatic switch off
10 minutes after last keypress
Display backlit LCD (on keypress)
Storage internal ring memory for 16 data sets
Interface IR interface for data transfer
Time real time clock und date
Calibration user and factory calibration
resetting to factory calibration possible
Dimensions 190 x 110 x 55 mm (LxWxH)
Weight approx. 455 g (incl. batteries)
Ambient conditions temperature: 5– 40°C
rel. humidity: 30–90% (non-condensing)
Waterproof as defined in IP 68 (1 hour at 0.1 meter); floatable instrument
*measured with standard solutions
To ensure maximum accuracy of test results, always use the reagent systems supplied by
the instrument manufacturer.