Getting started
CG-6 Operation Manual p/n 115370001 Rev A
Warm-up period: after you power up the CG-6 Autograv
TM
, it requires a 2 hour warm-
up time in order to reach the operating temperature.
Stabilization period: the instrument requires a 24 hour period after you power up to
stabilize.
Setting up the instrument for field operations: Once the steps have been performed
or attained, your CG-6 Autograv
TM
is ready for field use. For more details on the set up,
please refer to the next chapter entitled: Setting up your CG-6 Autograv
TM
.
Powering up the CG-6 Autograv
TM
The CG-6 Autograv
TM
can be powered either by:
• The 15V DC external power supply, or
Figure 2-8 Connecting the power supply to the CG-6 Autograv
TM
• by the two internal Smart Batteries supplied with the CG-6 Autograv
TM
.