Registers
Modbus RS485 Interface
Operating Guide
- 81 -
Registers
pH / Redox Input Service Alarm Configuration
10750 Get Channel 1 Service
Interval*
FLOAT Service Interval Value
*(Only available when
Service Reminder set to
yes)
0 to 999 Days
10820 Channel 2
10890 Channel 3
10752 Get Channel 1 Service Alarm
Date*
INT Service Alarm Date
Value
*(Only available when
Service Reminder set to
yes)
1 to 31 Day
10822 Channel 2
10892 Channel 3
10753 Get Channel 1 Service Alarm
Month*
INT Service Alarm Month
Value
*(Only available when
Service Reminder set to
yes)
1 to 12 Month
10823 Channel 2
10893 Channel 3
10754 Get Channel 1 Service Alarm
Year*
INT Service Alarm Year
Value
*(Only available when
Service Reminder set to
yes)
Max 2099 Year
10824 Channel 2
10894 Channel 3
Suspended Solids / Turbidity Service Alarm Configuration
10760 Get Channel 1 Service
Interval*
FLOAT Service Interval Value
*(Only available when
Service Reminder set to
yes)
0 to 999 Days
10830 Channel 2
10900 Channel 3
10762 Get Channel 1 Service Alarm
Date*
INT Service Alarm Date
Value
*(Only available when
Service Reminder set to
yes)
1 to 31 Day
10831 Channel 2
10902 Channel 3
10763 Get Channel 1 Service Alarm
Month*
INT Service Alarm Month
Value
*(Only available when
Service Reminder set to
yes)
1 to 12 Month
10833 Channel 2
10903 Channel 3
10764 Get Channel 1 Service Alarm
Year*
INT Service Alarm Year
Value
*(Only available when
Service Reminder set to
yes)
Max 2099 Year
10834 Channel 2
10904 Channel 3