LTSscale.com Page 27 of 28 REV F – 2018.04.09
LTS DR2100A/AN Indicator
Weight Calibration Numbers Standard and NTEP
This procedure can only be used after the initial calibration has been performed in the field and all numbers
in this section have been recorded. Entering calibration numbers can be performed in the event that an
electronics problem is encountered. This does not guarantee calibration is correct or within tolerance.
Scale should always be tested using known test weight before placing unit back into service.
This is not a substitute for a periodic calibration utilizing test weights.
1.43. Press 9091 then press ENTER Key.
9091
1.44. Record value.
LINEARITY STEP 1
0FFSET =
1.45. Press ENTER key. Record value.
LINEARITY STEP 1
ADD-WT =
1.46. Press ENTER key. Record value.
LINEARITY STEP 1
FACTOR =
1.47. Press ENTER key. Record value.
LINEARITY STEP 2
0FFSET =
1.48. Press ENTER key. Record value.
LINEARITY STEP 2
ADD-WT =
1.49. Press ENTER key. Record value.
LINEARITY STEP 2
FACTOR =
1.50. Press ENTER key. Record value.
LINEARITY STEP 3
0FFSET =
1.51. Press ENTER key. Record value.
LINEARITY STEP 3
ADD-WT =
1.52. Press ENTER key. Record value.
LINEARITY STEP 3
FACTOR =
1.53. Press ENTER key. Record value.
LINEARITY STEP 4
0FFSET =
1.54. Press ENTER key. Record value.
LINEARITY STEP 4
ADD-WT =
1.55. Press ENTER key. Record value.
LINEARITY STEP 4
FACTOR =
1.56. Press ENTER key. Record value.
SINGLE POINT CAL
FACTOR =
1.57. Press ENTER key. Record value.
SCALE ZERO
NUMBER =
1.58. Standard Unit, Recording complete.
NTEP unit, continue to Angle Calibration Numbers, next section.