Operating Manual V25/09.2019   
Ventus / Ventus-X / V200A
 
  Chapter 12 Configuration and Test  26 
12.2.6  Heating Parameters 
 
Figure 12: Heating Parameters 
 
Setpoint temperature:  Temperature to which the 
Ventus
 is heated  
default: 15°C 
Heater mode:  off:  heating is off  
auto:  automated control to setpoint temperature 
testmode: 
Ventus
 is heated to test setpoint temperature  
ctrl high:  heating off if control line on high, 
  otherwise: automatic control 
ctrl low:  heating off if control line on low 
  otherwise: automatic control 
  default: auto 
Capacity:  full:  full heating power  
alternating:  upper plate and lower plate alternatingly 
default: full 
Disable if overvoltage: 
the heating is disabled if overvoltage (28V) is detected  
default: yes 
Disable if undervoltage: 
the heating is disabled if undervoltage (20V) is detected  
default: no 
 
12.2.7  Wind Parameters 
 
Figure 13: Wind Parameters 
 
Min. speed:  starting speed of the wind sensor in m/s 
default: 0,1m/s 
Counts of measurements: 
Number of measured values used for determining the statistical values 
of average, minum value and maximum value 
default: 60 
Direction Offset:  Correction of the wind direction 
It is possible to have an offset added to the measured wind direction, 
in order to convert e.g. 0° (North) into 180° (South) 
default: 0° 
Measurement rate:  Update interval, adjustable in 250ms, 1 … 10 s 
default: 10s 
12.2.8  Pressure Parameters 
Pressure offset:  Absolute offset (for on-site adjustment) on the measured value