  Chapter 17 Fault Description  36 
17  Fault Description 
 
The device does not allow polling or 
does not respond 
-  Check supply voltage 
-  Check interface connection 
-  False device ID  check ID; devices are delivered 
with ID 1. 
Wind direction transmits incorrect 
values 
Device not correctly aligned  check alignment of 
device to North. 
Device transmits error value 28h 
(40d) 
Device is in initialization phase after start-up  
device delivers measurement values after ca. 10 
seconds 
Device transmits error value 50h 
(80d) 
Device is being operated above the specified 
measuring range 
Device transmits error value 51h 
(81d) 
Device is being operated below the specified 
measuring range 
Device transmits error value 55h 
(85d) for wind measurement 
Device unable to carry out valid measurement due to 
ambient conditions. 
There may be several causes for this: 
-  Device is being operated above the specified 
measuring range 
-  Very strong horizontal rain or snowfall 
- 
Ventus
 sensors are heavily soiled  clean sensor 
- 
Ventus
 sensors are iced up  check heating mode 
in configuration and verify function / connection of 
heating 
-  There are foreign bodies in the 
Ventus
 measuring 
section 
-  One of the 
Ventus
 sensors is faulty  return device 
to manufacturer for repair 
The quality of the wind 
measurement is not always 100% 
The device should always transmit 90 – 100% in 
normal operation. Values of up to 50% do not 
represent a general problem. 
When error value 55h (85d) is transmitted, this value 
is 0%. 
The device may be faulty if it permanently transmits 
values below 50%. 
Device transmits an error value not 
listed here 
There may be several reasons for this behavior  
contact the manufacturer’s technical support service. 
Minimum value of wind direction is 
greater than maximum value 
In the case of wind direction, the minimum / 
maximum value indicates the direction at which the 
minimum / maximum wind speed was measured.