IMPAC pyrometers series IN 5 plus
25
12 Maintenance
12.1 Safety
Attention duri
ng pyrometer services:
Should the pyrometer be integrated in a running machine process the machine has to be switched off and
secured against restart before servicing the pyrometer.
12.2 Service
The pyromet
er does not have any parts which require regular service, only the lens has to be kept clean.
The lens can be cleaned with a soft cloth in combination with alcohol (do not use acid solutions or dilution).
Also standard cloths for cleaning glasses or photo objectives can be used.
13 Trouble shooting
Before sending the pyrometer for repair, try to find the error and to solve the problem with the help of the
following list.
Temperature indication too low
Incorrect alignment of the pyrometer to the object
New correct alignment to achieve the max. temperature signal (see 6)
Measuring object smaller than spot size
check measuring distance, smallest spot size is at nominal measuring distance (see 7)
Measuring object is not always in the measuring spot of the pyrometer
Use max. value storage (see 9.2.1).
Emissivity set too high
Set lower correct emissivity corresponding to the material (see 9.1.1)
Len
s contaminated
Clean lens carefully (see 12.2)
Tempera
ture indication too high
Emissivity set too low
Set lower correct emissivity corresponding to the material (see 9.1.1)
The measurement is influenced by reflections of hot machine parts
Use mechanical construction to avoid the influence of the interfering radiation
Measuring errors
Indicated temperature is decreasing during the use of the pyrometer, contamination of the lens
Clean lens. Recommendation: use of air purge (see 12.2, 5.1)
Indicated temperature is decreasing during the use of the pyrometer, although the air purge unit is used.
Probably compressed air is not clean or air failed
Clean the lens and use clean, dry and oil free compressed air
Air contamination in the sighting path between pyrometer and object
Change position of the pyrometer with a clean sighting path (if necessary use a ratio pyrometer)
HF-interferences
Correct the connection of the cable shield (see 4)
Instrum
ent overheated
Use cooling jacket with air or water cooling (see 5.1)
Temperature Indication is fluctuating, probably caused by changing emissivity
Wrong pyrometer type, use of ratio pyrometer recommended
Laser targeting light
Laser targeting light fails
Instruments temperature above 50°C. Use cooling jacket (see 5.1)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LumaSense technologies IN 5 plus Series and is the answer not in the manual?

LumaSense technologies IN 5 plus Series Specifications

General IconGeneral
BrandLumaSense technologies
ModelIN 5 plus Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals