IGA 320/23 Operating Manual Data format UPP (Universal Pyrometer Protocol)
28
Description
Command
Parameters
Reading basic
temperature range:
AAmb
Output: XXXXYYYY (hex 8-digit, °C or °F)
XXXX = beginning of temp. range
YYYY = end of temp. range
Reading temperature
sub range:
AAme
Output: XXXXYYYY (hex 8-digit, °C or °F)
XXXX = beginning of temp. range (°C or °F)
YYYY = end of temp. range (°C or °F)
Setting of
temperature sub
range:
AAm1XXXXYY
YY
XXXX (hex 4-digit) beginning of temp. range (°C or °F)
YYYY (hex 4-digit) end of temp. range (°C or °F)
Address:
AAgaXX
XX = (00 ... 97)
00 ... 97 = regular device addresses
99 = Global address with response
98 = Global address without response
(only setting commands!)
Baud rate:
AAbrX
X=0...5 (decimal)
0 = 1200 Baud 2 = 4800 Baud 4 = 19200 Baud
1 = 2400 Baud 3 = 9600 Baud 5 = 38400 Baud
Temp. display
°C / °F
AAfhX
Output: X = 0: display in °C; X = 1: display in °F
Wait time:
AAtwXX
XX = 00 ... 99 (decimal, in bit time of the current baud
rate)
Internal
temperature:
AAgt
Output: XXX (dec. 000 ... 099 °C, 032 ... 210°F)
Max. internal
temperature:
AAtm
Output: XXX (dec. 000 ... 099, always in °C)
Limit switch SP1:
AAslXXXX
XXXX = set Limit switch, hex 4-digit, °C or °F
Mode
Limit switch SP1:
AAt1X
X = 0 Limit switch Off
X = 1 Limit switch close above
X = 2 Limit switch close below
Hysteresis
Limit switch SP1:
AAhlXX
XX = set hysteresis, hex 2-digit, °C or °F
Error status:
AAfs
Output: XX;
XX=00…FF (00 = no error)
(01…FF: error code for LumaSense service)
LED aiming light:
AAlaX
X = 0 switch off aiming light
X = 1 switch on aiming light
LED aiming light at
Power-on
AAlpX
X = 0 aiming light off at Power-on
X = 1 aiming light on at Power-on
Reading parameters:
AApa
Output 11-digit, decimal:
Digit 1 und 2 (10...99 or 00): Emissivity
Digit 3 (0 ... 6): Exposure time
Digit 4 (0 ... 8): Clear time max. / min. value storage
Digit 5 (0 ... 1): Analog output
Digit 6 und 7: (00 ... 98): Internal temperature
Digit 8 und 9 (00 ... 97): Address
Digit 10 (0 ... 6 or 8): Baud rate
Digit 11 always 0
Device type:
AAna
Output: “IGA 320/23” (16 ASCII-characters)
Serial number:
AAsn
Output: XXXXX (dec. 5-digit)

Other manuals for LumaSense 320 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LumaSense 320 Series and is the answer not in the manual?

LumaSense 320 Series Specifications

General IconGeneral
BrandLumaSense
Model320 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals