Holmium Surgical Laser
Operator's Manual
®
TM

Table of Contents

Questions and Answers

  • T
    Travis NicholsAug 17, 2025
    What to do if a notification appears on the Lumenis Medical Equipment control screen?
    • A
      Adam RodriguezAug 17, 2025
      If a notification appears on the control screen of your Lumenis Medical Equipment, first, clear the message by pressing the Acknowledge button and follow the suggested steps intended to clear the fault. If the same notification appears again, turn off the laser system for one minute and restart it to resume normal operation. If the notification continues to appear, record the error number and contact your local Lumenis service representative.
  • E
    Elizabeth WattsAug 20, 2025
    What to do if I get a Cooling System Notification on my Lumenis Medical Equipment?
    • L
      lwatkinsAug 20, 2025
      If you receive a cooling system notification on your Lumenis Medical Equipment, first press the Show Notifications icon to see the nature of the notification. Then, clear the message by pressing the Acknowledge button and pause operation for one minute to allow the system to cool. Resume normal operation; if the same notification appears again, turn off the laser system for five minutes and restart it. If the notification persists, record the error number and contact your local Lumenis service representative.
  • T
    thomas13Aug 22, 2025
    Why does my LUMENIS PULSE 120H system not turn on?
    • J
      Jennifer RojasAug 22, 2025
      If your Lumenis Medical Equipment system does not turn on, check the following: 1. Ensure the laser is plugged in. 2. Set the laser's main circuit breaker to the On (up) position. 3. Turn on the main electrical service power. 4. Try another outlet, or have the outlet professionally tested and repaired if necessary.
  • D
    Dana LongAug 26, 2025
    How to fix an Initialization Problem Screen on LUMENIS PULSE 120H?
    • A
      adam31Aug 26, 2025
      If your Lumenis Medical Equipment displays an Initialization Problem screen, start by writing down the error number. Then, press the Shutdown button. If the problem re-occurs, contact Lumenis service. Turn on the system with the keyswitch in the Off position. Export logs and send them to Lumenis service.
  • T
    tina86Aug 28, 2025
    What to do if the Fiber Expired on LUMENIS PULSE 120H?
    • C
      chapmanlisaAug 28, 2025
      If the Lumenis Medical Equipment fiber has expired, replace it with a new one and resume normal operation. If the problem persists, contact Lumenis Service.
  • N
    nicholas49Aug 30, 2025
    What to do if my Lumenis Medical Equipment shows 'Unrecognized Fiber'?
    • J
      Julian JonesAug 30, 2025
      If your Lumenis Medical Equipment shows 'Unrecognized Fiber', replace the fiber with a Lumenis compatible one and resume normal operation. If the problem persists, contact Lumenis Service.
  • B
    Brianna RobertsonSep 1, 2025
    What happens if the Fiber Exceeded Recommend # of Uses on my Lumenis Medical Equipment?
    • A
      Alexandra DanielSep 1, 2025
      If the Lumenis Medical Equipment fiber has exceeded the recommended number of uses, be aware of this and replace the optical fiber with a new one.
  • C
    Chad BarnettSep 4, 2025
    How to fix Fiber Burn Back on Lumenis Medical Equipment?
    • R
      Robin MartinSep 4, 2025
      If you experience fiber burn back with your Lumenis Medical Equipment, which may occur during prolonged procedures, especially when using higher power, renew the fiber tip by stripping and cleaving the fiber.
  • R
    Rachel BlackSep 6, 2025
    What to do if I hear 'Popping' or 'Tapping' sound from the Fiber Port of my Lumenis Medical Equipment?
    • T
      Terry NguyenSep 7, 2025
      If you are experiencing a 'Popping' or 'Tapping' sound coming from the fiber port of your Lumenis Medical Equipment, this is probably due to a malfunction of the fiber connector. Replace both the fiber and the debris shield.
  • T
    Teresa NguyenSep 9, 2025
    Why is there Inadequate or No Aiming Beam on my LUMENIS PULSE 120H Medical Equipment?
    • P
      pattonteresaSep 9, 2025
      If you observe inadequate or no aiming beam on your Lumenis Medical Equipment, adjust the aiming beam intensity. Also, replace the delivery system. Lower the intensity of the endoscopic camera light. Inspect and, if necessary, replace the debris shield. If the issue persists, contact your local Lumenis service representative.

Summary

System Description and Main Features

Holmium Laser Theory of Operation

Safety

Optical Hazards

Details risks associated with laser light exposure to eyes and skin.

Electrical Hazards

Outlines dangers related to electrical components and proper grounding.

Fire Hazards

Warns about flammable materials and potential ignition sources.

Protecting Non-Target Tissues

Provides guidelines to prevent accidental laser exposure to unintended areas.

Clinical Guide

Indications for Use

Lists the medical specialties and procedures for which the system is intended.

Contraindications

Lists conditions or situations where the laser should not be used.

Warnings and Precautions

Provides warnings and precautions applicable to surgical procedures with the system.

Preparing the System for Use

Connecting the Footswitch

Instructions for connecting the dual-pedal footswitch to the laser console.

Inserting the External Door Interlock Connector

Details on connecting the safety interlock to prevent unauthorized use.

Connecting the Delivery System

Procedures for properly connecting the fiber optic delivery system.

Normal Operation

Emergency Stop Button

Explains how to use the emergency stop button to immediately disable laser emission.

Powering on the System

Guides through the procedure of turning on the laser system.

Selecting the Treatment

Describes how to choose the correct specialty and preset for a procedure.

Starting Laser Treatment

Details the steps for activating the laser and initiating treatment.

Shutting Down the System

Procedures for safely powering down and disconnecting the laser system.

Advanced Operations

Moses Capability

Explains how to use and select modes for the Moses capability feature.

Troubleshooting and Maintenance

Handling Error Messages and Notifications

Explains how to view, acknowledge, and resolve system errors and notifications.

Troubleshooting

Provides solutions for common operational problems like screen issues or no aiming beam.

No Laser Emission

Addresses issues where the laser fails to emit energy.