Questions and Answers

  • J
    Jeffrey HowardSep 3, 2025
    How to troubleshoot inadequate aiming beam on Lumenis Medical Equipment?
    • E
      Edward ClaySep 4, 2025
      If you're experiencing an inadequate or no aiming beam with your Lumenis Medical Equipment, consider the following: * Adjust the aiming beam intensity. * Replace the delivery system if the optical fiber is defective. * Lower the intensity of the endoscopic camera light if it's too bright. * Inspect and replace the debris shield if it's blown. Note: If the debris shield requires replacement, also replace the delivery system optical fiber. * If the aiming beam is malfunctioning, contact your local Lumenis service representative.
  • J
    Jennifer GalvanSep 18, 2025
    What to do if Lumenis Medical Equipment shows 'Insert debris shield' message?
    • J
      Jonathan BarronSep 18, 2025
      If the Lumenis Medical Equipment control screen displays the “Insert debris shield” advisory message, it indicates that the debris shield is missing or is not properly inserted. Insert the debris shield as instructed in the “User Maintenance” section of this manual.
  • D
    David MartinezSep 16, 2025
    What to do if Lumenis Medical Equipment shows 'Check interlock' message?
    • A
      Amanda HicksSep 16, 2025
      If the Lumenis Medical Equipment control screen displays the “Check interlock” advisory message, it means that the interlock door is open, or the interlock plug is not properly inserted. Close the interlock door, or insert the interlock plug.
  • A
    Anne FrancisSep 14, 2025
    What to do if Lumenis Medical Equipment shows 'Attach footpedal' message?
    • N
      njamesSep 14, 2025
      If the Lumenis Medical Equipment control screen displays the “Attach footpedal” advisory message, it indicates that the footswitch is not properly connected to the laser. Connect the footswitch as instructed in the “Connection Instructions” section of this manual.
  • K
    Kyle AlvarezSep 13, 2025
    What to do if Lumenis Medical Equipment shows 'Attach fiber' message?
    • E
      Eric MillerSep 13, 2025
      If the Lumenis Medical Equipment control screen displays the “Attach fiber” advisory message, it means that the delivery system laser connector is not properly connected to the laser. Connect the delivery system as instructed in the “Connection Instructions” section of this manual.
  • M
    Michelle DelacruzSep 12, 2025
    What does '! or ! Service Required' mean on Lumenis Medical Equipment?
    • C
      candace92Sep 12, 2025
      When the Lumenis Medical Equipment displays '! or ! Service Required', it indicates that the laser is malfunctioning. Press or ready to clear the message. If the condition continues, turn off the laser for five seconds, then turn it back on. If the message reappears, record the advisory code from the summary screen and contact your local Lumenis service representative.
  • V
    Victoria McintyreSep 9, 2025
    What does '! or ! Service Soon' mean on Lumenis Medical Equipment?
    • V
      Valerie AndersonSep 9, 2025
      When the Lumenis Medical Equipment displays '! or ! Service Soon', it indicates that the laser is in Service Soon mode. This means that approximately five or fewer treatments remain before service is required. Schedule a service call with your local Lumenis service representative.
  • S
    Sandra TaylorSep 7, 2025
    What does '! ? 1.5J or ! Case Saver' mean on Lumenis Medical Equipment?
    • K
      Kristen PatelSep 8, 2025
      When the Lumenis Medical Equipment displays '! ? 1.5J or ! Case Saver', it indicates that the laser is in CaseSaver mode. In this mode, the maximum energy per pulse is reduced to 1.5 joules. When you finish the procedure, schedule a service call with your local Lumenis service representative.
  • A
    Andrew WalshSep 6, 2025
    What causes no laser power from the optical fiber in Lumenis Medical Equipment?
    • K
      Kelly ClarkSep 6, 2025
      If there is no laser power coming from the end of the optical fiber in your Lumenis Medical Equipment, the following could be the cause: * The delivery system optical fiber may be defective; replace the delivery system. * The debris shield may be blown; inspect and replace it. If the debris shield requires replacement, you should also replace your delivery system optical fiber. * The laser itself may be malfunctioning; contact your local Lumenis service representative.
  • A
    Amy BarreraAug 21, 2025
    What to do if Lumenis Medical Equipment displays '! or ! Overheating'?
    • U
      ubeardAug 21, 2025
      If the Lumenis Medical Equipment laser displays an '! or ! Overheating' message, it could be due to several reasons: * The laser may have been used at a high power for an extended period. Wait for the message to clear, press or ready, and continue using the laser, pausing occasionally to allow it to cool. Note: Do not turn off the laser, as leaving it on allows the internal cooling system to cool it down more quickly. * Airflow might be restricted. Ensure the laser is at least 50 centimeters (20 inches) away from walls, furniture, or other equipment. * The room temperature may be too high. Verify that the treatment room temperature is between 10° C and 30° C (50° F and 86° F).

Summary

Operation

Laser Console Basics

Explains how to turn on, restart, and turn off the laser console, including emergency shutdown procedures.

Control Screen Basics

Details the functionality of the control screen, including navigating tabs, changing display modes, and adjusting system settings.

Setting Treatment Values

Guides users on setting laser energy, pulse rate, understanding average power, and using CaseSaver mode.

Viewing the Summary Screen

Describes how to access and interpret treatment summaries, including laser usage statistics and individual treatment data.

Maintenance

Troubleshooting Guide

Provides systematic solutions for common laser system malfunctions, covering power, aiming beam, and connection issues.

User Maintenance

Covers essential user-level maintenance tasks, including cleaning the console and screen, and inspecting the debris shield.

Electrical Utilities

Details electrical requirements and configurations for the laser system, including voltage, current, and grounding specifications.

Specifications

Lists the technical specifications of the laser system, covering treatment beam, aiming beam, input power, and cooling requirements.

Safety and Regulatory

Laser Safety Eyewear

Details laser safety eyewear requirements, hazard distance calculations, and optical density standards for eye protection.

Additional Safety Considerations

Covers important safety practices, including effects of spot size changes, material handling, reflections, and smoke evacuation.

Protecting nontarget tissues

Offers guidance on preventing accidental laser exposure to unintended tissues and ensuring safe operation.

Regulatory Compliance

Summarizes compliance with FDA and CE standards, and explains key safety features like emergency stops and interlocks.

Indications for Use

Indications for Use Specific to the Ho:YAG Wavelength

Lists medical specialties where the Ho:YAG wavelength is used and provides essential general warnings for its application.

Urology indications

Details urological applications, contraindications, warnings, and clinical parameters for Ho:YAG wavelength procedures.

E.N.T. surgery indications

Lists applications for endonasal/sinus surgery, including procedures like turbinectomy and polypectomy using the Ho:YAG wavelength.

General surgery indications

Lists applications for various open, laparoscopic, and endoscopic general surgery procedures using the Ho:YAG wavelength.

LUMENIS VersaPulse PowerSuite 20 Specifications

General IconGeneral
BrandLUMENIS
ModelVersaPulse PowerSuite 20
CategoryMedical Equipment
LanguageEnglish

Related product manuals