User Interface Commands
FOT Lab Kit User’s Manual 4-15
MU Report/Set Measurement Reporting Interval
The MU command specifies the interval between measurement
reports. Another way of looking at it is that 1 / MU = reporting
frequency.
Information Request:
Syntax: <Esc>MU ? <CRLF>
Response: MU = interval
Example: MU = 4.5 S
Parameter Modification:
Syntax: <Esc>MU = interval <CRLF>
Range: C or 'Continuous
Continuous mode reporting sets the reporting
interval to the minimum practical value i.e. the
sample interval times the number of active
channels. Since the sample rate is 4 Hz the
sample interval is then 250 milliseconds and
the reporting interval will be 250mS * channels.
Examples:
If 4 channels are active MU=C would yield 250mS * 4 = 1.0 S.
If 2 channels are active MU=C would yield 250mS * 2 = 0.5 S.
0.25 - 600 S Interval may be specified in seconds
1 - 10 M Interval may also be specified in minutes
The minimum interval corresponds to the sample rate of 0.25
Seconds (4Hz).
The maximum interval is limited by the width of the timer (16 bits).
Examples: <ESC> MU = C <CRLF>
<ESC> MU = 25 <CRLF>
<ESC> MU = IM <CRLF>