User Interface Commands
4-16 FOT Lab Kit User’s Manual
PS Active Input Channels
The PS (Probe Select) command specifies which input channels are
to be activated.
It is recommended that unused inputs be disabled to provide the
optimum number of samples per report for the active channels.
The number of active channels has a direct effect on the
interpretation of the MU command parameter value 'C' i.e.
continuous reporting.
Information Request:
Syntax: <Esc>PS ? <CRLF>
Response: PS = active channel list
Example: PS = 1,2,3,4
Parameter Modification:
Syntax: <Esc>PS = channel list <CRLF>
Range: Any combination of installed channels may
be specified, in any order with the channel
numbers separated by commas. Additional
spaces may be used but are ignored if
present.
Examples: PS = 1, 2, 3, 4
PS = 2, 4
PS = 3, 1, 2