47
8.3. Continuous corneal curvature / refractive power measurement
mode (K&R mode)
This is the mode for carrying out the corneal curvature measurement and refractive power measurement
continuously.
- K&R mode selection: Set so that the measurement mode indicator section on the screen becomes
“K&R” mode.
8.3.1. Manual measurement mode
! NOTE
The manual measurement mode has a minimum limit set to allow measurement even in
unusual situations. Therefore, data errors may occur depending on the user's skill level. In
general, automatic measurement mode is recommended.
1 (Eye height adjustment), (measurement location and focusing) process is carried out just like the
8.1.1 refractive power measurement mode.
2 Measurement
–
Press on the measurement button.
–
Measurement is carried out continuously when measurement button is pressed on continually.
–
Measurement result is indicated on the monitor when measurement is completed.
–
The most recent measurement result is indicated when continuous measurement is taking place.
[K&R mode indicator screen]