When using an FDA or CE-cleared (as appropriate) disinfecting agent, carefully follow the
instructions provided the manufacturer of the product.
For low-level disinfection(normally), the patient contact parts may be wiped with any of the
following low-level disinfectants Methods to disinfect to LRK-7000 are as below:
- Dry heat
- Mechanical cleaning with disposable wipe / sterile gauze
- Wipe with gauze soaked in alcohol or chemicals like hydrogen peroxide and Merthiolate
- Soaking in chnicals like 70% isopropyl alcohol, 1:1000 Merthiolate, 3% hydrogen peroxide and 1:10
diluted house hold bleach (sodium hypochlorite)
Solution
Manufacturer
Cleaner/
Disinfectant
Active ingredient
Cleared/Approved
for use in
Alkazyme
Alkapharm
Cleaner
Proteolytic enzyme,
Quat, Ammonia
Europe
Klenzyme
Steis/Calgon Corp.
Cleaner
Enzymes
USA & Europe
- For high-level disinfection (if needed), the patient contact parts may be wipe using one of
the following disinfection agents:
Solution
Manufacturer
Cleaner/
Disinfectant
Active
Cleared/Approved
Cidex
OPA
Advanced
Sterilization
Product
Disinfectant
Orthophtalade-
hyde
USA & Europe