86
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 6100-4-3
3 Vrms
150 kHz to 80
MHz
6 Vrms 150 kHz
to 80 MHz out
side ISM bandsa
3 V/m
80 MHz to 2.7 GHz
3 Vrms
150 kHz to 80 M
Hz
6 Vrms 150 kHz
to 80 MHz outside
ISM bandsa
3 V/m
Portable and mobile RF communications
equipment should be used no closer to
any part of the Model LRK-7000, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d=[
3.5
𝑉
1
]
𝑃
d=[
3.5
𝐸
1
]
𝑃 80MHz to 800MHZ
d=[
3.5
𝐸
1
]
𝑃 80MHz to 2.7GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters(m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, a should
be less than the compliance level in
each frequency range
b
Interference may occur in the
vicinity of equipment marked with
the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.