Operation Manual MA 25/MA 25e
MA 27/MA 27e
To avoid the risk of electric shock, this equipment must only
be connected to the medical power supply originally
delivered by MAICO. Using another power supply can also
lead to electrical damage on the device.
Prevent cable breakage: cables must not be bent or
buckled.
2.7 Device Control
The user of the device should perform a subjective device check once a week
according ISO 8253-1. See section 6.7 for a checklist.
For annual calibration please see sections 2.5 and 3.1
2.8 Electromagnetic Compatibility (EMC)
This device is suitable in hospital environments except for
near active HF surgical equipment and RF shielded rooms of
systems for magnetic resonance imaging, where the intensity
of electromagnetic disturbance is high.
The device fulfills the relevant EMC requirements. Avoid
unnecessary exposure to electromagnetic fields, e.g. from mobile
phones etc. If the device is used adjacent to other equipment it
must be observed that no mutual disturbance appears.
Use of this device adjacent to or stacked with other equipment
should be avoided because it could result in improper
operation. If such use is necessary, this instrument and the
other equipment should be observed to verify that they are
operating normally.
Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of this equipment
could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and
result in improper operation.
The list of accessories, transducers and cables can be found
in the section 6.5 of this operation manual.
Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should be used
no closer than 30 cm (12 inches) to any part of the MA 25/MA
25e/MA 27/MA 27e, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this
equipment could result in improper operation.