6.5 Electromagnetic Compatibility (EMC)
ESSENTIAL PERFORMANCE for this device is defined by the manufacturer as:
This device does not have an ESSENTIAL PERFORMANCE.
Absence or loss of ESSENTIAL PERFORMANCE cannot lead to any unacceptable
immediate risk. Final diagnosis shall always be based on clinical knowledge.
This instrument is in compliance with IEC60601-1-2:2014, emission class B group 1.
NOTE: There are no deviations from the collateral standard and allowances uses.
NOTE: All necessary instruction for maintaining compliance with regard to EMC can be
found in the general maintenance section in this instruction. No further steps required.
To ensure compliance with the EMC requirements as specified in IEC 60601-1-2, it is
essential to use only the accessories listed in the following table. Conformance to the
EMC requirements as specified in IEC 60601-1-2 is ensured if the cable types and
cable lengths are as specified.
ITEM
MANUFACTURER
MODEL
CABLE
LENGTH
[M]
SCREENED
(YES/NO)
Audiometric
Headphones
Radioear
DD45
2.0
Yes
(MA 27 only)
Audiometric
Headphones
Radioear
DD65 v2
2.0
No
Patient
response
switch
Radioear
APS3
2.0
Yes
Power Supply
(Wall plug)
UE / Fuhua
UES18LCPU-
050200SPA
1.5
No
USB cable
Type A/B
Sanibel
8011241
2.0
Yes