Operation Manual MA 33
8100092 Rev. 9
56
19/08/2019
Columns shown in Speech Audiometry table (Figure 76 and Table 13):
Figure 76
Table 13 Explanation of Display Text
Display text
Information
SRT
dB HL, level at which speech reception threshold is obtained
SAT
dB HL, level at which speech awareness threshold is obtained test
ear during an SRT/SAT test
Mask (1)
dB HL, level of masking noise presented
%
percent correct score obtained for a word recognition (discrimination) test.
Signal
dB HL, level at which a word list was presented to the test ear for a
word recognition (discrimination) task
Mask (2)
dB HL, level of masking noise presented to the non test ear during
a word recognition (discrimination) task
Condition/Word List
lists transducer, ear, and lists what was presented
Notes
the user may type comments into this field
MCL
dB HL, value obtained for most comfortable listening level to speech
UCL
dB HL, value obtained for uncomfortable listening level to speech
SNR loss
Value calculated from the results of the QuickSIN test.
If WAVE file is selected, a speech list is displayed on the screen (Figure 77).
1
2
Figure 77