Operation Manual
touchTymp MI 26
and
MI 36
Version
Environmental
conditions:
+15 °C to +35 °C /
+ 59 °F to +95 °F
Relative humidity 30 % to 90 % (non-condensing)
Air pressure 98 kPa to 104 kPa
3
Maximum altitude: 2000 m / 6561 ft above sea level
Warm up time: 10 minutes (including boot up time)
0 °C to + 50 °C / 32 °F to +122 °F
Humidity 10 to 95 % (non-condensing)
-20 °C to + 50 °C / -4 °F to +122 °F
Humidity 10 % to 95 % (non-condensing)
300 mm x 345 mm x 148 mm
11.81 in x 13.58 in x 5.83 in
204 mm x 25 mm x 26 mm
8.03 in x 0.98 in x 1.02 in
104 mm x 36 mm x 24 mm / 4.09 in x 1.42 in x 0.94 in
Tubing: 2175 mm / 85.63 in
10.4 in full color display with high bright white LED back-light
Chinese, English, French, German, Italian, Polish, Russian,
Spanish, Turkish
External / USB out, USB in, USB out, power socket, Contra
headphone jack, probe connector
USB 1.1 / Ethernet (not implemented)
USB; the system can not be operated from a PC.
Using MAICO Sessions together with the OtoAccess Database,
Noah or a Practice Management Software via BDT/GDT-
interface (only for Germany, Austria and Switzerland), data can
be transferred and saved on the PC.
Thermal printer
(configuration
dependent):
110 mm width, 20 m length
To be printed on paper roll:
200 Tympanograms
87 Tympanograms with Acoustic Reflexes for
both ears
4 s (one Tympanogram) to 12 s
(Tympanogram with Acoustic Reflexes for both ears)
Environmental conditions during operating according IEC 60645-1.
NOTE: Reference equivalent threshold sound pressure levels may differ significantly with ambient
pressures outside the above range. Therefore recalibration around the normal ambient pressure at the
site of the user should be undertaken in those circumstances where the calibration site and the user
site do not share similar ambient conditions.