EasyManuals Logo

Maico touchTymp MI 26 User Manual

Maico touchTymp MI 26
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
Operation Manual
touchTymp MI 26
and
MI 36
Version
8511780 Rev. 4
30/06/2021
Calibration Values: Bone Conductor Radioear B81
Coupler ISO 389-3, ANSI S3.6
Frequency
[Hz]
Reference equivalent
threshold
force level for tone
Max level
ISO 389-3
[dB] (re 1µN)
Tone
[dB HL]
250
67.0
45
500
58.0
70
750
48.5
75
1000
42.5
80
1500
36.5
85
2000
31.0
80
3000
30.0
75
4000
35.5
70
6000
40.0
50
8000
40.0
40
White Noise
42.5
65

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Maico touchTymp MI 26 and is the answer not in the manual?

Maico touchTymp MI 26 Specifications

General IconGeneral
BrandMaico
ModeltouchTymp MI 26
CategoryDiagnostic Equipment
LanguageEnglish