Warning
Properly place defibrillation pads
Do not place electrodes near the generator of
an internal pacemaker. Always apply
electrodes to flat areas of skin. Avoid
application over folds of skin such as those
underneath the breast or on obese patients.
Excessive hair, poor adhesion, or air under
electrode may produce burns.
2.3.5 ECG Misinterpretation
Warning
Properly place defibrillation pads
Improperly placed pads may produce incorrect
analysis and an inappropriate shock or no
shock advisory decision.
Warning
Handling or transporting the patient during
ECG analysis can cause incorrect or delayed
diagnosis. Follow all instructions in the Users
Manual.
Warning
Cardiac pacemakers may affec
t rhythm
Patient pacemakers may reduce the sensitivity
of the AED7000 analysis and errors in
detecting shockable rhythms.
Warning
Radio frequency (RF) interference
Do not operate the AED7000 in conjunction
with electro-cautery or diathermy equipment.
Any equipment that emits strong radio