6 Maintaining the Automatic External Defibrillator
The maintaining, cleaning and Troubleshooting of the AED7000 will be described
in this section:
Maintaining
Cleaning and disinfecting
Troubleshooting
6.1 Maintenance
Maintenance is required for the optimal operation of the AED7000. Maintenance
should be performed periodically; with weekly or monthly inspections
recommended.
The frequency of inspection:
It is recommended that the AED7000 is inspected once a week or twice a month,
but if use of the AED7000 is more frequent, additional inspections should be
performed.
The table below describes a general rule to perform the
inspection:
Frequency of use (n / month)
inspection
1 / month
Once a week
2 ~ 3 / month
Twice a week
Infrequent using(1 ~ 2 a year)
Once a month
Caution
If the AED7000 should be used more than 4 times a
month, we recommend that the AED7000 should be
inspected by authorized service personnel at least
twice a year.