25
4.3 Self-test function description in detail
Daily-test: On commencement, the self-testing LED indicator will turn on; and
automatically turn off on completion of self test. The self-testing result will be
displayed in battery LCD, as follows:
a. If the battery is low the following icon will appear:
b. If the battery is in normal condition the following icon will appear:
Monthly-test: On commencement, the self-testing LED indicator will turn on;
when finished, the self-testing LED indicator turns off. And the test result will
appear in LCD
a. If there is a fault with the AED7000 the following icon will appear:
b. If the AED7000 is in optimal working order the following icon will appear:
4.4 Actions required during AED Self-testing
1) The AED7000 will only function under normal operating condition if the
following icon is shown:
2) Monthly-test when Battery changed
If the monthly self-test function is not required, please install the new battery
within 10 seconds following removal of the old battery
If the monthly self-test function is required, please install the new battery 20
seconds later following removal of the old battery. The monthly self-test will
commence one minute later after installation of the new battery.