Output
Energy
(J)
Patient
Impedan
ce
(Ω)
I
p1
(Amps)
I
p2
(Amps)
Phase1
(ms)
Phase2
(ms)
Interval
(ms)
200 25 46 34.0 5.5 3.3 0.7
50 23 17.0 10.9 6.6 0.7
75 15.3 11.3 16.3 8.9 0.7
100 11.5 8.5 19.7 9.6 0.7
125 9.2 6.8 20.5 11.2 0.7
150
7.6
5.8 19.8 10 0.7
175
5.2 19.8 10 0.7
Pediatric Energy Reducer:Energy Attenuation Factor
50%, at 25 Ω( ,
25% at 50Ωand 75Ω (
20%, at 100Ωand 125Ω
15%, at 150Ωand 175Ω
7.4 Battery
Category Specification
Non-rechargeable 12V d.c. 2.8Ah
Capacity 100 discharges at 200 Joules or 120
discharges at 150 Joules
℃
℃
Note