32
Deliver a shock:
To deliver a shock, the several steps below must be obeyed:
Make sure the indicator of shock button is flashing, which means the AED7000 is
properly charged.
Before pressing the shock button, make sure no one is touching the body of the
patient.
Press the shock button.
Note
If the shock button is pressed, shock indicators and
analyzing indicator will stop flickering while the
flickering of Caution indicator will continue.
Prior to pressing the shock button, if the AED7000 detects that
the pads are not firmly placed on the chest of the patient the
following alert will occur
Shock cancelled, check the electrodes.
If the operator does not press the shock button, the AED7000 will
automatically begin internal discharging procedure in 30 seconds.
Time out, analysis restart.
5.8 Cardiopulmonary resuscitation procedure (CPR)
The AED7000 will enter cardiopulmonary resuscitation procedure, when the
condition below occurs:
Note
During the CPR procedure, the four indicators: Check