33
Pad, Caution, Analyzing and Shock will not flicker.
No shockable heart rhythm is detected
During the heart rhythm analysis period, if the heart rhythm is not considered to be
a shockable rhythm by the AED7000, a cardiopulmonary resuscitation procedure
(CPR) will start.
No shock advised
During the period of charging, if the AED7000 detects that the shockable rhythm
has changed to a normal one, the AED7000 will stop current procedure and switch
to CPR procedure.
1. The operator can touch the body of the patient and check their
conditions.
Note
During the cardiopulmonary resuscitation period, the
AED7000 will ignore the heart rhythm analyzed result
and the improper electrode connection.
If the patient is not breathing and has no pulse, a cardiopulmonary resuscitation
should be performed on the patient immediately.
Note
The CPR time is fixed at 120 seconds; the AED7000 will
prompt the rest time every 30 seconds.
2. The voice prompts of rest time are below:
Patient may be touched. Carry on CPR. Alternately give
30 chest compressions and 2 mouth-to-mouth breaths.
Now give 30 times chest compressions…...Now give 2
mouth-to-mouth breaths……
Four Cycles