External Electrical
Connections
No external devices are attached to the unit
Internally powered equipment with
defibrillator-proof BF type patient applied
part (as per definition of IEC 60601-1
standard)
7.3 Defibrillator
Output Specification
Energy Sequence (user configurable)
Biphasic Truncated
exponential
150, 150, 200J
Charge Time from “Shock Advised
10 sec. to 150J
15 sec. to 200 J
Analysis Time
9 sec.
From start of analysis to
shock ready:
from power up
Less than 20 seconds
after 6 shocks
Less than 25 seconds
Audible Prompts
Visual Prompts
LED prompts
Controls
Two buttons - On/Off, Shock
Output Energy Accuracy
±15% into any impedance
from 25 to 175Ω
Waveform details
The table below provides details of the biphasic truncated exponential waveform
delivered by the AED7000 (set to 200J) when connected to resistive loads of 25
through 175 Ohms. The waveforms are characterized by typical values for peak
current (Ip), duration of the first output phase, and duration of the second output
phase. Values are within 10%.