Solution:
If the electrode pads are dry, damaged or out-of-date, replace the
current pads with new ones and recommence.
Ensure that the electrode pads have been properly adhered to the
patient. If not attached correctly, wipe off any water or moisture
from the chest and shave the hair from chest area, then press the
pads firmly on the chest of the patient.
Ensure the two pads do not touch each other.
If the pads connector is inadequately connected to the socket of
the AED7000, push the connector firmly into the socket.
6.3.2 Troubleshooting of analysis
Problem description:
The analysis is interrupted during the analysis period, and the voice prompt of
“Check the electrode” is broadcast.
Solution:
Shut down the AED7000 and check the pads placed on the
patient, if they are not properly connected, press the pads firmly
on the chest of the patient, then start the AED7000 again.
If the patient has moved during the analyzing period, shut down
the AED7000 and immobilize the patient if possible.
6.3.3 Troubleshooting of charging
Problem description:
The charging is stopped during the charging period, and the voice prompt of
“Charging cancelled: check the electrode” is broadcast.
Check the electrodes.