Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5% U
T
(>95%
dip in U
T
) for
0.5 cycle
40% U
T
(60%
dip in U
T
) for 5
cycles
70% U
T
(30%
dip in U
T
) for 25
cycles
<5% U
T
(>95 %
dip in U
T
) for 5
sec
Not applicable
Not applicable
Not applicable
Not applicable
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE. U
T
is the AC mains voltage prior to application of the test level.
7.6.3 Electromagnetic immunity (IEC 60601-1-2 Table 203)
The AED7000 is intended for use in the electromagnetic environment specified
below. The customer or the user of the AED7000 should ensure that it is used in
such an environment.
Immunity
test
IEC 60601
test level
Complian
ce level
Electromagnetic environment –
guidance
Radiated
RF IEC
61000-4-
3
10V/m 80
MHz to
2.5 GHz
20V/m
Portable and mobile RF
communications equipment should be
used no closer to any part of the
AED7000, including cables, than the
recommended separation distance