3
CAUTION! Use a licensed electrician
for all wiring.
WARNING: Failure to disinfect the
dental operating light between patients
could expose user/patient to cross
contamination and bio-burden/bio-
contamination.
WARNING: To avoid risk of electric
shock, this equipment must be
connected only to supply mains with
protective earth.
WARNING: No modication of this
equipment is allowed.
WARNING: Power cords and their
associated parts cannot be substituted
without increase risk of electric shock
or re. Use manufacturer-approved
replacement parts only! Power
cords must be installed by qualied
personnel. Make sure all service
loops, strain reliefs, and cord guards
are in place and that line, neutral
and ground wires are secured.
WARNING: The light is not to be used
in rooms where an explosion hazard
exists.
WARNING: Rotational stops are
designed for a maximum force limit of
30 lbs. If rotational stops are broken
immediately take the equipment out of
use and have the equipment serviced.
Operation of damaged equipment
could result in injury to the operator
and/or patient.
Storage Conditions:
-55°C to +50°C
10% to 90% Relative Humidity
The pre-installation must be performed according to the requirements in our ‘Pre-installation Instructions’.
As manufacturers of electro-medical products we can assume responsibility for safety-related performance of
the equipment only if maintenance, repair and modications are carried out only by us or agencies we have
authorized for this purpose, and if components affecting safe operation of the unit that may be needed are
replaced with original parts.
We suggest that you request a certicate showing the nature and extent of the work performed, from those
who carry out such work, and specify that the certicate show any changes in rated parameters or working
ranges, as well as the date, the name of the rm and a signature.
SAFETY NOTES
WARNING: Failure to install set
screws & roll pin per Installation
Instructions could result in injury or
damage to equipment.
WARNING: To avoid risk of burn injuries
do not touch the light reector during
operation.
WARNING: To avoid risk of injury, avoid
contact with sharp edges on dental
lights.
CAUTION: Only authorized service
technicians should attempt to service
this equipment. Use of other than
authorized technicians will void the
warranty.
WARNING: To avoid risk of injury, avoid
contact with sharp edges on dental
lights.