4.2 Starting up
1. Snap the EMMA Airway Adapter into EMMA. It will click into place when properly seated.
2. Press the Power On button.
3. When EMMA is ready the ETCO
2
value is zero.
The audible alarm sound may be checked by detaching the Airway Adapter to generate a “Check
Adapter” alarm.
When EMMA is ready the end-tidal carbon dioxide (ETCO
2
) display indicates 0 and the Respiratory
Display indicates “- -“.
If the ETCO
2
display shows a non-zero value, ensure that there has not been an accumulation of CO
2
between EMMA and the EMMA Airway Adapter by removing and reattaching the EMMA Airway Adapter.
If the ETCO
2
still displays a non-zero value after this procedure, perform a Zeroing procedure as
described in chapter 7.4 prior to using EMMA with a patient.
4.3 Switching off
EMMA switches off automatically 15 Sec after that the EMMA Airway Adapter is removed or 2 min after
a No Breath condition is detected and the Alarm Silence button is pressed.
13(32)

Other manuals for Masimo EMMA

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Masimo EMMA and is the answer not in the manual?

Masimo EMMA Specifications

General IconGeneral
ParametersEtCO2, Respiration Rate
Measurement Range0-150 mmHg
DisplayLCD
Battery LifeUp to 4 hours of continuous operation
Weight200 grams
TechnologyInfrared spectroscopy
EtCO2 Range0 to 150 mmHg
Respiration Rate Range0-99 breaths/min
Sampling MethodMainstream

Related product manuals