8.2 Electromagnetic Compatibility (EMC)
Guidance and Masimo’s declaration – electromagnetic emissions –
for EMMA
EMMA is intended for use in the electromagnetic environment specified below. The customer or the user
of EMMA should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
EMMA uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B EMMA is suitable for use in all establishments including
domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
26(32)