EasyManua.ls Logo

Masimo EMMA - Cautions; Notes

Masimo EMMA
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WARNING! Replace batteries immediately when the Battery Status indicator starts blinking.
Remaining battery time depends on battery type and other circumstances and cannot be
reliably predicted. The remaining lifetime for lithium batteries may be significantly less than 30
minutes when the Battery Status Indicator starts blinking.
WARNING! Lithium batteries may present a fire or chemical burn hazard if mistreated. Do not
disassemble, heat above 100°C (212°F) or incinerate. Dispose of used cell promptly. Keep
away from children.
WARNING! Use only Alkaline or Energizer Ultimate Lithium L92 batteries. Use of other Lithium
batteries may present a risk of fire or explosion.
2.2 Cautions
CAUTION! If EMMA is used in a manner other than that for which it was intended, unpredictable
behavior could result.
CAUTION! The EMMA Airway Adapters are non-sterile devices. Do not autoclave the devices as this
will damage them.
CAUTION! Never sterilize or immerse EMMA in liquid.
CAUTION! Do not operate EMMA at ambient temperatures less than -C (23°F) or greater than
50°C (122°F).
CAUTION! Remove batteries if EMMA is not likely to be used for a period of time longer than 90 days.
2.3 Notes
NOTE! Throughout this User’s Manual:
EMMA Airway Adapter refers to both Airway Adapter Adult/Pediatric and Airway Adapter Infant if not
otherwise mentioned.
NOTE! Always carry spare batteries in the EMMA pouch.
6(32)

Other manuals for Masimo EMMA

Related product manuals