1. 19,999 (main)/1,999 (secondary) dual LCD display
Remarks: This accuracy is the measurement standard. In
DUT jack, the meter-specific probe should be used if
required. Measurement with probe may be influenced by
external environment. To avoid inaccurate measurements,
please keep away from strong magnetic sources.
4. Functional Characteristics Description
2. LCR measures with automatic identification and automatic
measurement range
3. L\C\R single measurement selection
4. Resistance measurement in DCR mode
5. D/Q/θ/ESR display on the secondary display
6. Series or parallel measurement modes are selectable
7. Comparison function under single measurement
8. Test frequency 100/120/1k/10k/100k is selectable in
AC mode
9. The selection feature for measured components with the
same series
10. Battery power display, auto power off if meter is not
operated for 5 minutes.
11. With infrared transmission interface, the meter is secure
(supporting hot plug). With special software, the meter is
easily managed.
12. See table 1-3 for measurement accuracy and scope
Table 1 Resistance measurement scope
Measurement
Measuring
Frequency
Range
Resolution
Accuracy
100Hz/120
200.00Ω 0.01Ω
1.0%+5d
100Hz/120
2.0000kΩ 0.1Ω
0.3%+5d
100Hz/120
20.000kΩ
1Ω
0.3%+5d
100Hz/120
200.00kΩ 0.01kΩ
0.5%+5d
100Hz/120
2.0000MΩ 0.1kΩ
1.0%+5d
100Hz/120
20.000MΩ
1kΩ
1.0%+5d
100Hz/120
200.00MΩ 0.1MΩ
2.0%+5d
1kHz
1kHz
1kHz
1kHz
1kHz
1kHz
1kHz
200.00Ω
0.001Ω
1.0%+5d
0.01Ω
0.3%+5d
0.1Ω
0.3%+5d
1Ω
0.3%+5d
0.01kΩ
0.5%+5d
0.1kΩ
1.0%+5d
1kΩ
2.0%+5d
0.1MΩ
5.0%+5d
1kHz
2.0000kΩ
20.000kΩ
200.00kΩ
2.0000MΩ
20.000MΩ
200.00MΩ
20.000Ω
10kHz
10kHz
10kHz
10kHz
10kHz
100kHz
100kHz
100kHz
100kHz
200.00Ω
0.001Ω
1.0%+5d
0.01Ω
0.5%+5d
0.1Ω
0.3%+5d
1Ω
0.5%+5d
0.01kΩ
1.0%+5d
1.0%+5d
1.0%+5d
1.0%+5d
2.0000kΩ
20.000kΩ
200.00kΩ
20.000Ω
200.00Ω
2.0000kΩ
20.000kΩ
20.000Ω 0.001Ω
0.01Ω
0.1Ω
1Ω
2.0%+5d
Rs/Rp
Note: This accuracy is the measurement standard when
D <0.1, if D> 0.1, it should be multiplied by the extraction
of a root of 1 + D².