Maxim Integrated Page 37 of 52
WSpO
2
Algorithm Mode 1
(MAX32664GWEC)
10x the measured value of SpO
2
.
SpO
2
calculation progress (%) - only in one-shot mode.
lowSNRflag: A flag showingwing low SNR .
Motion flag: A flag showing highexcessve motion.
isWSpo2 State: Algorithm reported state:
0: Adjustment state
1: Computation state
2: Successful state
3: Timeout stateIsWSpo2Calculated: Algorithm reported status.
Calibration/estimation mode status.
0: No Signal
1: In Progress
2: Successful
3: Bad Signal
4: Implied Motion Detected
5: Failure
6: Calibration Done
Calibration mode progress. Value ranges from 0 to 100. 100 meaning calibration
is complete.
Heart rate (bpm). LSB = 0.1 bpm.
Systolic BP value. Only valid once progress = 100.
Diastolic BP value. Only valid once progress = 100.
SpO
2
value (70% - 100%): 16-bit, LSB = 0.1%
r: 16-bit LSB = 0.1 SpO2 r value.
HRaboveResting flag: a flag showing non-resting heart rate (>95 bpm)