easyjet pro - manual of use and maintenance
EN
05.2 Instructions for use
1 Insert the front terminal of the polisher (sterile) (Fig.3 - Ref.A) so that it ts closely;
2 Connect Easyjet pro to the turbine of the dental unit (Fig.3 - Ref.B);
3 Adjust the water ow rate on your dental unit in order to obtain a steady stream of water. It is easier to
adjust the ow rate without powder;
4 Unscrew the cap of the power container (Fig.4 - Ref.A);
5 Pour powder into the container, making sure that the level does not reach the nozzles inside the
container;
WARNING: It is advisable to use the measuring cap of the pack of prophylaxis powder, lling it to
the correct level for the Easyjet pro.
WARNING: It is advisable not to shake the Easyjet pro after lling it with powder, as this could
clog the nozzles.
6 Screw the cap of the powder container back on, without tightening it excessively (Fig.4 - Ref.B);
WARNING: Remove any residues of powder from the thread of the cap before screwing it on.
7 Press the turbine pedal to carry out the treatment;
WARNING: Use the suction cannula to evacuate the mixture of air, bicarbonate and water from
the patient’s mouth.
8 Carry out the treatment using small circular movements. Direct the nozzle only towards the enamel
of the teeth, from the gingival sulcus towards the incisal edge of the tooth, keeping at a distance of
between 3 and 5 mm.
9 Upon completion, polish all the surfaces of the teeth, setting the water ow rate to the maximum;
WARNING: The stream of air, bicarbonate and water will continue for several seconds after
removing your foot from the control pedal. Take these few extra seconds into account when nishing
the treatment.
10 After the treatment, the treated teeth will be clean and the dental cuticle will have been completely
eliminated. It will take the proteins contained in the saliva between two and three hours to form it
again. For this lapse of time, the teeth will no longer have their natural protection against stains. The
patient must be warned not to smoke and not to eat or drink anything that might stain the teeth (tea,
coffee, etc.).
IMPORTANT NOTE: Front terminal of the polisher clogged (no powder coming out).
1 Disconnect the front terminal of the polisher from the main body of the Easyjet pro (Fig.3 - Ref.C).
2 Use the cleaning needle supplied with the device to free the channel from any residues of powder
(Fig.2 - Ref.C), carrying out the movements indicated (Fig.5). Blow compressed air into the central
hole of the terminal from both ends, using the syringe.
WARNING: Clean the channel of the nozzle using only the cleaning needle supplied with
the device (Fig.2 - Ref.D).
3 Also clean the front connecting part of the casing of the Easyjet pro thoroughly. If necessary, remove
any residues of powder by blowing compressed air onto them with the syringe.
4 Connect the front terminal of the polisher to the Easyjet pro (Fig.3 - Ref.A).