20
35.0-400.0mL /h(60d/mL IV set)
Auto-calculating the speed by the bolus volume
11 levels are available
225mmHg~975mmHg,11 level
1:225mmHg
2:300mmHg
3:375mmHg
4:450mmHg
5:525mmHg
6:600mmHg
7:675mmHg
8:750mmHg
9:825mmHg
10:900mmHg
11:975mmHg
Near End, Infusion End, Occlusion Alarm, Low Battery, Battery
Empty, No Battery, No Power Supply, Door open, Air Bubble, No
Drop Sensor, No Drop Signal, Drop Error, Reminder Alarm, and
alarms during self-test and running.
Repeat alarming: If there is still alarm after mute alarm sound, it
will alarm again in 2 minutes.
Event recording: can store and playback 1000 events maximum
Alarm volume setting: 5 levels of alarm voice are available.
Power supply switching: When AC/DC power supply is cut off,
the infusion automatically switches
to internal battery supply.
Barcode scanning: Input the patient information by barcode
scanning.
Ambient temperature: 5℃-40℃ Relative humidity: 15%-95%,
nocondensation Atmospheric pressure: 70.0 kPa-106.0 kPa
Ambient temperature: -20℃-60℃ Relative humidity:
10%-98%,
no condensation Atmospheric pressure: 22.0.0 kPa-107.4.0 kPa