31
Note!
The graduate to inspect the infusion precision,is just used for general maintenance.
The special materials for disassembling should refer to the requirements in the
disassembling and installation chapter.
The safety project inspection needs specialized testing equipments.After disassembling
and installing the device, the safety of the pump must be tested.
5.1 Alarm information
During normal operation, the pump generate visible and audible alarm to remind users to deal
with the alarms in time.
The alarm can be categorized into clinical alarm and technical alarm.Clinical alarm is
generated according to the operation state during the general using of the device.It is not
equipment failure but to remind users to deal with the device working in time,such as when
aninfusion is finished or a reminder alarm.A clinical alarm is sounded along with concrete
wording of specific information on the screen.
A technical alarm is caused by abnormal data inside the device during normal use of the
device. It can be an equipment failure or not, showing as ERR-XXXX.
5.1.1 The Alarm Levels
Give out a beep at the time of
every alarm.
The yellow indicator
illuminates.
Give out three beeps at intervals
of 15 seconds.
Theyellow indicator
flashing.
Give out a series of beeps at
intervals of 15 seconds.
The red indicator flashing.
Give out two short beeps.
The corresponding
indicator flashing.
5.1.2 The Clinical Alarms and Troubleshooting
No external AC/DC power
supply is connected.
Immediately connect the
AC power supply or the
external DC power supply.
No internal battery or the
internal battery operates