32
The internal battery is
running critically low.
Immediately connect an
AC power supply or an
external DC power supply.
The infusion will finish
within 3 minutes.
Wait until the infusion
finishes.
1. The tubing is occluded.
2. Low occlusion level
has been set during
high viscosity.
Click [STOP] to clear the
alarm, troubleshoot, and
then continue the infusion.
Door is open during
infusion.
Click [STOP] to clear the
alarm. Check whether the
pump door has been
closed.
1. Air bubble in the
infusion line.
2. The flatten tube is fixed
inside the air bubble
detector.
Click [STOP] to clear the
alarm. Check the tubing to
release the air bubble.
The VTBI or the infusion
time set finishes.
Click [STOP] to clear the
alarm.
Operation forgotten (no
operation on any key for 2
minutes after the IV set
loaded.
Click any key to clear the
alarm.
5.1.3 Technology alarms and troubleshooting