Installation SI PCF User Manual
Ref: DOC03148 Issue 1 Friday, 06 January 2017 Page 28 of 81
Connecting the SI PCF
Note: Medisafe (UK) Ltd. recommends this product be installed & commissioned by Medisafe (UK)
Ltd (or suitably qualified personnel).
1. Unpack the machine and position it
within 1m of the mains power supply,
drain, and mains water supplies.
2. Remove any protective packaging and
loose items from inside the wash tank.
3. Identify the grey plastic, convoluted
drain hose supplied with the unit.
Connect the small end of the hose to
the ¾” drain hose tail at the lower rear
of the unit.
4. Secure the drain hose to the machine
using the supplied jubilee clip.
5. Connect the large end of the drain
hose to the site drain, cutting to length
as required.
6. Identify a stainless steel water hose
supplied with the unit. Ensure the seal
in the hose end is in place and screw
the elbow connector onto the cold-
water inlet at the lower rear of the unit.
7. Connect the straight end of the hose
to a suitable cold water supply.
8. Repeat for the hot water supply.
9. Repeat for the RO water if this has
been made available on site.
10. Turn on the water supplies and check
for leaks.
11. Connect the machine to the electrical
mains supply (see section 2-7).
12. Carry out the commissioning
procedure defined within the SI PCF
Service Manual.