Re-Wash Cycle
As a water and time saving feature, Program 2
is a re-wash program. This allows the wash
water and chemical to be reused. At the end
of the wash, the user is allowed to open the lid,
add new instruments, and repeat the wash
stage using the same water up to ten times.
Note: if the water temperature exceeds the set
point by more than 10°C the cycle will be
aborted (see section 3-12).
At the end of a wash stage press Yes on the
screen to reuse the water again.
Unload the instruments and load more
instruments for cleaning.
Temperature of the water within
the unit can reach 60°C
(140°F), take care when
opening lid and reloading the
instruments.
Press Reload Complete on the screen to
begin another wash stage
To end the re-wash process, once the wash
stage is complete press No on the screen to
drain the water and end the cycle.