Operation SI PCF User Manual
Ref: DOC03148 Issue 1 Friday, 06 January 2017 Page 40 of 81
Selecting a program
The SI PCF is installed with six pre-programmed wash cycles during manufacture. When starting a
cycle, the machine will use the currently selected program. To change the selected program, prior to
initiating a cycle:
1. Use the scroll arrows to cycle
through to the Wash program icon
2. Change the selected program with the
+ and – buttons
Standard Program Parameters
Parameters Units
Program Number
P1 P2 P3 P4 P5 P6
Pre-wash Time Minutes 3 0 3 3 3 3
Wash Chemical 1 ml 360 360 360 360 360 360
Wash Temperature ºC 40 40 40 40 40 40
Degas Minutes 3 3 3 3 3 3
Wash Time Minutes 5 10 20 5 10 20
Rinse Temperature ºC 60 60 60 60 60 60
Rinse Time Minutes 2 2 2 2 2 2
Disinfection Temperature ºC 0 0 0 90 90 90
Disinfection Time Minutes 1 0 1 1 1 1
Typical Total Cycle Time
Minutes 40 10* 50 55 60 70
*Per Re-wash, see section 3-11-10