15
Calibrating the
Medit i700
Then follow the steps from the Calibration Wizard. Always turn the dial ANTICLOCKISE while
calibrating.
The calibration process is essential in order to produce precise 3D models. It is highly
recommended that you perform the calibration process periodically.
1. Calibration is required when:
The quality of the 3D model is not reliable or accurate as compared to previous results.
Environmental conditions such as temperature have changed.
Calibration period has expired:
You can set the calibration period in Menu > Settings > Calibration Period(Days).
The default calibration period is set to 14 days.
2. Precautions when performing calibration
The calibration panel is a delicate component and you should not touch it directly.
Should there be issues with the calibration process, check the panel first. If the
calibration panel has been contaminated, please contact your service provider.