Cleaning and Sterilizing
Medit i700
SUMMARY
1. Handpiece: Wipe down with safety cap ON
(Max strength: 60-70% alcohol/Vol.)
2. Tip Mirror: Water rinse and soft gauze with 70% alcohol
3. Seal in steripouch DRY!
4. Heat Setting: *Important* Plastic Cycle!
Sterilize at 121(249.8°F) for 30 minutes
DRY CYCLE for 15 minutes!
__________________________________________________________________________
1. Reusable tip
The reusable tip is the part that is inserted into the patient’s mouth during scanning. The tip is
reusable for a limited number of times. The tip needs to be cleaned and sterilized between
patients to avoid cross-contamination.