The tip should be cleaned manually using a disinfecting solution. After cleaning and
disinfecting, inspect the mirror inside the tip to ensure that there are no stains or
smudges.
Repeat the cleaning and disinfection process if necessary. Carefully dry the mirror using
a paper towel.
Insert the tip into a paper sterilization pouch and seal it, making sure that it is airtight.
Use either a self-adhesive or heat-sealed pouch.
Sterilize the wrapped tip in an autoclave with the following conditions:
» Sterilize for 30 minutes at 121°C (249.8°F) at gravity type and dry for 15 minutes.
» Sterilize for 4 minutes at 134°C (273.2°F) at pre-vacuum type and dry for 20 minutes.
Use an autoclave program that dries the wrapped tip before opening the autoclave.
Scanner tips can be re-sterilized up to 100 times and must thereafter be disposed of as
described in the disposal section.
2. Disinfection and Sterilization
Clean the tip immediately after use with soap water and a brush. We recommend using
a mild dishwashing liquid. Make sure the mirror of the tip is completely clean and stain-
free after cleaning. If the mirror appears stained or foggy, repeat the cleaning process
and rinse thoroughly with water. Dry the mirror carefully with a paper towel.
Disinfect the tip using Wavicide-01 for 45 to 60 minutes. Please refer to the Wavicide-01
Solution instruction manual for proper use.
After 45 to 60 minutes, remove the tip from the disinfectant and rinse thoroughly.
Use a sterilized and non-abrasive cloth to gently dry the mirror and the tip.
Caution
The mirror found in the tip is a delicate optical component that should be handled with
care to ensure optimal scan quality. Be careful not to scratch or smudge it as any
damage or blemishes may affect the data acquired.
Make sure to always wrap the tip before autoclaving. If you autoclave an exposed tip, it
will cause stains on the mirror which cannot be removed. Check the autoclave manual
for more information.
New tips need to be cleaned and sterilized/autoclaved before their first use.
Medit is not responsible for any damage including distortion, blackening etc.
3. Mirror
The presence of impurities or smudges on the tip mirror may lead to poor scan quality and an
overall poor scanning experience. In such a situation, you should clean the mirror following the
steps below: