Warning Troubleshooting Guide
Indicates the error information for Hemo-Force®, these include: Lo,
Hi, GE and F.
Alarm
Codes
Fault Type Description Troubleshooting
LO Low Pressure 1. Leg is less than
40mmHg pressure;
foot pressure is less
than 120mmHg after
2 consecutive cycles.
2. Air hose is
disconnected from
garments.
1. Check connection
between garment and
air hose.
2. Check for leaks in the
garments or air hose set
and replace if faulty.
3. Turn the pump off and
restart.
4. Refer to a professional
for service if problem
persists.
HI High Pressure 1. Air hose is kinked.
2. Leg pressure is more
than 40mmHg; foot
pressure is more than
120 mmHg after 2
consecutive cycles.
1. Check air hose set for
any kinks.
2. Refer to a professional
for service if problem
persists.
GE Garment Error 1. Garment is not
connected to air hose
during start up.
2. Garments are not
fully deflated during
each restart.
3. The initialization
failed due to a
possible blockage in
air hose/garment
connectors.
1. Attach garments to
air hose and restart.
2. Ensure garments are fully
deflated before restart.
3. Check connectors for any
blockage and restart.
4. Refer to a professional
for service if problem
persists.
F
System Failure 1. Loose air hose
connection on both
ports at pump.
2. Pressure is lower than
set pressure after 10
consecutive inflations.
3. Pressure is greater
than set pressure
after two inflations.
4. Electrical malfunction.
1. Check air hoses
connections.
2. Turn the pump off and
restart.
3. Refer to a professional
for service if problem
persists.