Model No. Hemo-Force®
Item Name Intermittent DVT Pump
Ref No. MDS600INT2
Size 7.54”x 5.12”x7.95”
Weight 5 lbs (2.8 Kg) - without battery
Pressure Range Calf/Thigh Garment: 40 mmHg +10/-5mmHg (5.3kPa
+1.3kPa/-0.7 kPa)
Foot Garment: 120 mmHg +10/-5 mmHg (16 kPa +1.3/-
0.7 kPa)
Input Rating AC 100-240V, 50/60Hz, 0.6~0.4A
Power Cable 15ft, non-shielding, AC powered (USA)
Fuse Rating T1AH 250V
Classification Class II, Type BF
Not AP or APG type
Ingress Protection IP22
Operation Humidity 30–75%
Operation Temperature 15–35°C
Operation Atmospheric
Pressure Range
700 hPa to 1060 hPa
Mode of Operation Continuous
Applied Part Garment
Battery Pack Model.
(optional)
MDS600BAT (14.4 Vdc, 2200 mAh / 31.68 Wh
Lithium-ion rechargeable battery pack)