19
Electromagnetic Compatibility
That portable RF communications equipment can affect medical
electrical equipment. We recommend a safety distance no closer
than 30 cm (12 inches) to any part of the DVT Pump, including
cables and at least 3.3 feet (1 meter) for sensitive equipment.
Medical electrical equipment needs special precautions regarding EMC
and needs to be installed according to the EMC information provided.
Manufacturer’s Declaration – Electromagnetic Emissions
The Hemo-Force® is intended for use in the electromagnetic environment (for professional
healthcare) specified below. The customer or the user of the Hemo-Force® should ensure
that it is used in such an environment.
Emission test Compliance Electromagnetic
environment-guidance
(for professional healthcare
environment)
RF emissions
CISPR 11
Group 1 The Hemo-Force® uses RF
energy only for its internal
function. Therefore, its RF
emissions are very low and
are not likely to cause any
interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A The Hemo-Force® is suitable
for use in all establishments
other than domestic and those
directly connected to the public
low-voltage power supply
network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable (120V)
Class A (230V)
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable (120V)
Compliance (230V)
CAUTION: This equipment is not intended for use in residential
environments and may not provide adequate protection to radio
reception in such environments.